In the pulling sequence…

Model 720 Needle Pipette Puller

is the ideal instrument for the busy laboratory where a number of pipettes are required. This unit is easy-to-use, versatile and a reliable vertical pipette puller with excellent repeatability.

By setting or adjusting values for the various parameters in the pulling sequence, you can vary the length, taper and tip to achieve an almost endless variety of shapes and sizes.

Model 720 will repeatedly pull micropipettes of the configuration for which it has been set. This is possible because of the ability to control the parameters in the pulling sequence and the vertical slide mechanism. Parameter controls include a heater control knob and digital display adjustable to 0.1 resolution, DC regulated solenoid control and optical switch with adjustable flags to set the length of gravity fall. These controllable parameters coupled with heater configuration allow a wide variety of pipettes to be pulled.

Small fluctuations in heater temperature can significantly affect the shape of a pipette during a pulling sequence. To prevent this, Model 720 is designed to be less susceptible to line voltage changes. Heater voltage is divided into
20 units with a resolution of 0.1 units.

Another factor which can adversely affect pipette shape is residual heat in the heating elements. When using platinum alloy filaments, Model 720 helps eliminate this problem in several ways. First, the heater goes off the instant the solenoid is activated, and since the unit is of vertical design, any residual heat rises away from the tips. Second, platinum/iridium filaments have an extremely low thermal mass and cool very rapidly. Third, the tip of the pipette is pulled below the heating elements where there is no residual heat. These solutions are so effective there is no need for air jet cooling.

Model 720 Advanced Features

720

 


Size: 9½" W x 14½" H x 11" D

Weight: 35 lbs (total shipping)

Power Requirements: 100/120/220/240 VAC,
50-60 Hz via rear panel entry module with appropriate power cord.

Digital Display: Three digits, resolution of tenths.
0-19.5 units with resolution of 0.1 units.

Solenoid Current: 0-5 amp DC regulation.

Optical Switch: Fixed location with adjustable flags.

Fuses: 4 amp, 1.5 amp (100/120 VAC);
3.15 amp, 1.6 amp (220/240 VAC)

 

Printable Copy