Kopf Electrode Angle Calibrator provides a
quick and reliable method for accurate stereotaxic placement of electrodes at any angle, including right and compound angles. Unit consist of an aluminum base plate and a stainless steel calibrated A/P bar, lateral slide and a rotational vertical post mounted on a zeroing point. The lateral slide has a 0.1 mm vernier attached to a moveable slide block. When the block is set at zero, the zeroing point of the slide is at ear bar zero.
To use the calibrator, A/P, lateral and vertical
coordinates are determined, the angle selected
and the lateral slide moved to the proper co-ordinate. The manipulator is placed on the A/P bar; the selected angle is set. The manipulator
is then adjusted until the electrode tip is at the proper coordinate on the vertical scale. |
|
The A/P and vertical position of the manipulator are then noted and the vertical slide of the manipulator retracted enough to clear the head
of the animal when placed on the A/P bar of the stereotaxic frame. Since the noted A/P position
is at A/P zero, it is necessary to add or subtract from the dimension to be at the proper A/P coordinate. Advance the vertical slide of the manipulator down to previously noted vertical reading to arrive at the final position.
By removing the rotational vertical post and setting
the lateral slide block at zero, the angle calibrator
can also be used as an A/P zeroing bar.
Model 1455 is used with the 1460 & 1760-SB
series manipulators (18.70 mm A/P bar).
Model 1755 is used with the 1760 & 1460-LB